The MiNi-HAC from APREL is a Hearing Aid Compatibility Automated Test System that can test devices operating up to 6 GHz. It is ideal for use as a primary compliance instrument for Hearing Aid Compatibility or as a design/development tool for electromagnetic near-field emission evaluation of antennas and wireless devices. This system is equipped with E-Field and H-Field HAC probes, each with two standard frequency calibrations. It complies with ANSI/IEEE C63.19 standards, ensuring accuracy and reliability in electromagnetic testing. The test system utilizes automated X, Y, Z precision probe movement system with a 5-axis laboratory-grade robot.
The MiNi-HAC comes with a user-friendly GUI that allows for easy setup and data retrieval. The software supports signal generator (no additional equipment needed), full FFT (Fast Furrier Transform) analysis, multiple octave settings (user defined or standard) and tone generation (comes with ITU P50). It includes full graphic package for visualization and manipulation of measured data. This test system has 4 input channels and uses RS232 and USB interfaces for packet data communication. It includes an active twin axis T-coil probe, standard frequency validation dipoles, and a telephone magnetic field simulator (TMFS) for reference checks. This system requires a bench area of less than 6 sq. feet.